| Keterangan Lainnya |
"- Operating frequency range: from 1 to 10,0 MHz
- Range of measured time intervals (duration of scanning): from 6 to 1000 µs
- Velocity range: 1000 – 9999 m / s
- Error of measurement of time intervals: not exceed ± 0,025 µs
- Maximum permissible error of measurement of the amplitudes of the signals at the receiver input in the range from 0 to 110 dB: not exceed ± 0,5 dB
- Testing gain range: 125 dB
- Averaging over the quantity of starts: from 1 to 16
- Range of variation of temporal sensitivity adjustment (TVG): 40 dB
- Number of control points TVG: 16
- Duration of the excitation pulse to the load: from 0,0 to 0,5 µs
- Deviation of the amplitudes of input signals in the range from 10 to 100% of the screen height not more than: 1 dB
- Setting a delay in the prism of the probe: from 0 to 15 µs
- Detection of signals: positive half-wave, radio mode
- Standards:
• EN 12668
• ASTM E1324
• EN 55011
• EN61000-6-2:2001
• EN 61010-1:2014 (EN 61010-1:2010)
• EN 60529:2014
• EN ISO 16810:2016
• SOU NNEC 027:2014
• SOU NNEC 032:2014
• SOU NNEC 038:2017
- Dimensions (W*H*L): 80*162*38 mm
- Weight, not more: 250 g (without batteries)"
|